Control range
- Temperature range from 0 to 120°C Humidity range from 30 to 95% RH

Performance
- Microprocessor PID control / Temperature calibration / Automatic tuning.
- Rapid humidity control is achieved by adapting the humidification tank interior to the chamber..
- Air-tight door closure with perfect sealed packing.
- Heat-resistant silicone packing completely blocks high heat leakage from the inner chamber.
Convenience
- Two different sized models available to suit any users’ needs.
- Corrosion resistant stainless steel chamber.
- Cable port for introduction of external probes or wires.
- 50mm diameter cable port comes standard on the left side of the unit.
- Additional 50mm or 80mm diameter cable ports, custom ordered. (optional)
- Inner glass door with entry ports for handling samples without affecting inner chamber temperature and humidity. (optional)
- Casters for easy mobility during installation or relocation.
- User friendly LCD controller.
- Colorful interactive touch type LCD controller..
- Set point temperature and present temperature are clearly shown together on display.
- Sufficient programmable control support.
- Operating progress can be shown in graph mode for better data understanding.
- Automatic start/stop can be programmed for unattended operations.
- Display indicates abnormal temperature changes or power failures.
- Selective bilingual language support.
(Korean / English or English / Chinese, selectable only when placing an order) - Temperature and other operating parameters can be recorded using a six channel digital recorder.(optional)
- Computer interface.
- RS-485 communication port (optional) enables connection up to maximum 32 units. Simultaneous monitoring and operating of multiple units is an available feature.
- Program patterns are simple to set with the use of our software.
- Easy to check status of real-time operations. Graph data supports easy understanding.
- Storage and documentation of programs.
- Data with saving capabilities in excel and text format.
- Storage of program information, backup of value settings, recovery, and temperature values.
- RS-232 communication port as standard offer.

| Description | No | |
| Programmable pattern capacity | 100 | |
| Pattern repeat time | 999 | |
| Part repeat time | 255 | |
| Max. segments per a pattern | 100 | |
| Available max. segments* | 2000 | |
| Programmable process time per a segment | 0 to 99 hr. 59 min. | |
* Even though the max. segments per pattern are 100, the total available segments are up to 6000 not 10000, therefore arrangement of the number of segment and patterns will need to be done properly.
Stability test ICH guideline Q1A (R2)
- Operating progress visible through graph mode for better data analyzation.
- Cyclic defrost system, designed to work excellently for the long period stability test operation. (※Temperature and humidity changes may occur during automatic defrost.)
- Maintenance of water supply are easily performed with a front water tank.
- Water level check indicated by water level bar in the tank.
- Water supply during operation is also available for long term test purposes.
- Direct water system provides user great convenience when performing long-term test. (optional)
Photostability test ICH guidelineQ1B (Option 2)
- Built in UV and fluorescent lamps for uniformed luminescence.
- Equipped with movable UV and VIS sensors .
- Independent lamp controller
- An indicator for Total usage time of UV and fluorescent lamps.
- Convenient set function of cumulative dosage and lamp Of
Safety
- Built in door lock with a key.
- Automatic shut off after overheat alarm.
- Low and empty water level alarm.
- Power supply leakage breaker.
- Over current protection.
- Open door alarm.
| Models | TH-ICH-300 | TH-ICH-800 |
||||
| Chamber volume (L / cu ft) | 300 / 10.6 | 760 / 26.8 | ||||
| Controller | PID control (LCD Type) | |||||
| Temperature | Range (°C / °F) | without humidity (°C / °F) | -5 to 85 / 23 to 185 without Light 0 to 85 / 32 to 185 with Light |
|||
| with humidity (°C / °F) | +15 to 85 / 59 to 185 with Light | |||||
| Fluctuation 1) (±°C / °F) at 40°C / 60%RH | 0.3 / 0.55 | 0.3 / 0.55 | ||||
| Variation 1) (±°C / °F) at 40℃ / 60%RH | 0.5 / 0.9 | 0.7 / 1.3 | ||||
| Time | Heating | 25min (20℃ → 85℃) | 30min (20℃ → 85℃) | |||
| Cooling | 35min (20℃ → -5℃) | 40min (20℃ → -5℃) | ||||
| Humidity | Range (%RH) | 20 to 85 ※ See page 231 for more detailed humidity range check. |
||||
| Fluctuation2) (±%RH) at 60%RH / 40°C | 3 | |||||
| Variation2)(±%RH) at 60%RH / 40°C | 5 | |||||
| Light source (ICH Q1B option 2) | Cool white3) | 6,000 lux | 6,500 lux | |||
| Ultraviolet-A3) | 4.5 W/㎡ | 5 W/㎡ | ||||
| Light uniformity | VIS. : ±10%UVA : ±15% | VIS. : ±15%UVA : ±15% | ||||
| Typical time required to reach ICH Recommendation of 1.2 million lux-hr |
≒ 200Hr | ≒ 184Hr | ||||
| Typical time required to reach ICH recommendation of 200 W-hr/m2 |
≒ 50Hr | ≒ 40Hr | ||||
| Refrigeration | System | Air – cooled | ||||
| Refrigerant | R- 404A | |||||
| Water tank | Capacity (L / cu ft) | 15 / 0.53 | ||||
| Water quality | pH6.2 ~ 7.2, Electrical conductivity 20㎲ / cm to below | |||||
| No. of wire shelves (standard / max.) | 2 / 8 | 2 / 17 | ||||
| Distance of between shelves (mm / inch) | 60 / 2.4 | 60 / 2.4 | ||||
| Max. load per shelf (kg / lbs) | 25 / 55.1 | 40 / 88.2 | ||||
| Permitted total load (kg / lbs) | 75 / 165.3 | 100 / 220.5 | ||||
| Dimension (W×D×H) | Interior (mm / inch) | 750×650×650 / 29.5×25.6×25.6 | 750×880×1160 / 29.5×34.6×45.7 | |||
| Exterior (mm / inch) | 1130×985×1520 / 44.5×38.8×59.8 | 1220×1215×1965 / 48×47.8×77.4 | ||||
| Net weight (kg / lbs) | 280±10 / 617±22 | 400±10 / 882±22 | ||||
| Cable port | Basically one hole provided / Ø 50mm (2”), Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2”) |
|||||
| Interface port | Basically RS-232C, [RS-485 : optional] | |||||
| Electrical requirements | AC230V, 1ph, 60Hz | 15.4 A | 24.5 A | |||
| Cat.No. | AAHC1001K | AAHC1011K | ||||
| AC230V, 1ph, 50Hz | 14.7 A | 23.5 A | ||||
| AC230V, 1ph, 50Hz | 14.7 A | 23.5 A | ||||
| Cat.No. | AAHC1002K | AAHC1012K | ||||
4320 Polypropylene 0.5ml Graduated Flat Top Microcentrifuge Tube, Amber Bio Plas
NH310, Portable Colorimeter 3nh
NS800, Spectrophotometer 3nh
3970223- 2' Purifier Vertical Clean Bench with Airflow Monitor- Labconco
02-070MGXL-12 - Multi-Gen Adapter (Required for Multi-Gen 7XL Probes)- PRO Scientific
3970221- 2' Purifier Vertical Clean Bench with UV Light and Protection Panel- Labconco
NR200, Precision Colorimeter 3nh
NR110, Precision Colorimeter 3nh
9120010-801 Micrometer reticule Labomed
NBL 12001P -Nimbus® Precision Balances- Adam Equipment
9120010-802 Micrometer reticule, scale 10mm divided in 100 parts, horizontal scale with locking circlip Labomed
850071 Environmental Quality SD Card Logger Sper Scientific
PAA480BO01-Anti-MYO Rabbit Polyclonal Antibody- Cloude Clone
4318 Polypropylene 0.5ml Graduated Flat Top Microcentrifuge Tube,Blue Bio Plas
V0020-F1 Hydrophobic filter, pk/5 ( 1 each included) Accuris
V1002 Handheld vacuum controller with single channel adapters Accuris
VV0020-BOT Graduated, PC collection bottle (included) Accuris
4316 Polypropylene 0.5ml Graduated Flat Top Microcentrifuge Tube, Red Bio Plas
Q620 WashLab760 Laboratory Bottle Washer Kimadi
V1002-8 Optional eight channel adapter, for use with 200ul pipette tips Accuris
V1002-S8 Optional eight channel adapter, fine tip stainless steel Accuris
WashLab790 Laboratory Bottle Washer Kimadi
911015119999 D2012 Plus High Speed Personal Micro-Centrifuge Scilogex
840003 Digital Anemometer Thermometer Sper Scientific 