- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results.
- Temperature range 5 °C (32 ºF) above ambient temperature up to 300 °C (572 °F).
- MS controller with serveral timer functions.
Controller timer functions: delayed ON, delayed OFF, temperature dependent delayed OFF
- Digital temperature setting with an accuracy of one degree.
- Adjustable fan speed.
- Front ventilation flap slide and rear exhaust duct Ø 50 mm (1.97 inch).
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm.
- RS 422 interface for communication software APT-COM™ DataControlSystem, or switch over to printer output with RS 232 / RS 422 interface converter
- Units up to 115 liters are stackable.
- 2 chrome-plated racks included.
- BINDER test certificate.
| Specifications | |
| Exterior dimensions | |
| Width (mm/inch) | 1234 / 48.6 |
| Height (inclusive feet/castors) (mm/inch) | 1022 / 40.2 |
| Depth (mm/inch) | 765 / 30.1 |
| plus door handle, I-panel and exhaust duct (mm/inch) | 105 / 4.1 |
| Wall clearance rear (mm/inch) | 100 / 3.9 |
| Wall clearance side (mm/inch) | 160 / 6.3 |
| Exhaust duct outer-Ø (mm/inch) | 52 / 2.1 |
| Steam space volume (l/cu.ft.) | 498 / 17.6 |
| Number of doors | 2 |
| Interior dimensions | |
| Width (mm/inch) | 1000 / 39.4 |
| Height (mm/inch) | 800 / 31.5 |
| Depth (mm/inch) | 500 / 19.7 |
| Interior volume (l/cu.ft.) | 400 / 14.3 |
| Racks, chrome-plated (number standard/max.) | 2 / 10 |
| Load per rack (kg/lbs.) | 35 / 77 |
| Permitted total load (kg/lbs.) | 90 / 199 |
| Weight of the unit (empty) (kg/lbs.) | 145 / 320 |
| Temperature data | |
| Temperature range, 5 °C (41 °F) above ambient up to (°C / °F) | 300 / 572 |
| Temperature variation 1) | |
| at 70 °C (± °C) | 1 |
| at 150 °C (± °C) | 2.5 |
| at 300 °C (± °C) | 4.8 |
| Temperature fluctuation (± °C) | 0.3 |
| Heating up time 2) | |
| to 70 °C (Min.) | 18 |
| to 150 °C (Min.) | 35 |
| to 250 °C (Min.) | 60 |
| Recov. time after door was opened for 30 sec. 2) | |
| at 70 °C (Min.) | 2 |
| at 150 °C (Min.) | 17 |
| at 300 °C (Min.) | 21 |
| Air change data | |
| Air change | |
| at 70 °C (x/h) | 17 |
| at 150 °C (x/h) | 18 |
| at 300 °C (x/h) | 16 |
| Electrical data | |
| Housing protection acc. to EN 60529 | IP 20 |
| Nominal voltage (±10 %) 50 / 60 Hz (V) | 400 |
| Nominal power (kW) | 3.4 |
| Energy consumption | |
| at 70 °C (W) | 520 |
| at 150 °C (W) | 1200 |
| at 300 °C (W) | 2340 |
0091F Assorted Fluorescent Polypropylene Reversible 1.5ml And 2.0ml Microcentrifuge Tube Rack, 96 Places (Pack Of 5) Bio Plas
0090 Natural Polypropylene Reversible 1.5ml And 2.0ml Microcentrifuge Tube Rack, 96 Places (Pack Of 5) Bio Plas
0092 Blue Polypropylene Reversible 1.5ml And 2.0ml Microcentrifuge Tube Rack, 96 Places (Pack Of 5) Bio Plas
911015119999 D2012 Plus High Speed Personal Micro-Centrifuge Scilogex
0095F Fluorescent Pink Polypropylene Reversible 1.5ml And 2.0ml Microcentrifuge Tube Rack, 96 Places (Pack Of 5) Bio Plas
840003 Digital Anemometer Thermometer Sper Scientific
Q620 WashLab760 Laboratory Bottle Washer Kimadi
0094F Fluorescent Orange Polypropylene Reversible 1.5ml And 2.0ml Microcentrifuge Tube Rack, 96 Places (Pack Of 5) Bio Plas 